4.7 Calibration
4.7.1 Calibrate the active sensor with control (ACT wCON)
Calibration possibilities:
- After pressing the ENTER key, the automatic calibration begins. The device will read the 0% Load value and
display WARM UP. After that the 100% Load value will be read and the device will display WARM UP. After the
WARM UP is finished the device will display SAFE?. Press ENTER to confirm or ESC to reject.
4.7.2 Calibrate the active sensor without control (ACT nCON
)
Calibration:
- measure 0% Load and measure 100% Load
- measure 0% Load and enter the difference [100% Load in V – 0% Load in V]
- enter 0% Load in V and enter difference [100% Load in V – 0% Load in V]
Press key
or and choose between 0% Load or Sensitivity
0% Load - Relieve sensor from load
or
Sensitivity - Enter the value fort the sensitivity in V
Press Key
or
and choose between 100% Load or Sensitivity
100% Load - Calibrate with 100% Load
or
Sensitivity - Enter the value for the sensitivity in V
SAFE - Accept calibration values
4.7.3 Calibrate the passive sensor with control (ACT wCON)
Calibration:
The device will read the 0% Load value and display WARM UP. After that the 100% Load value will be read and the
device will display WARM UP. After the WARM UP is finished the device will display SAFE?. Press ENTER to
confirm or ESC to reject.
4.7.4 Calibrate the passive sensor without control (ACT nCON)
Calibration:
- measure 0% Load and measure 100% Load
- measure 0% Load and enter 100% Load in mV/V
- enter 0% Load in mV/V and enter 100% Load in mV/V
Press key
or and choose between 0% Load or Sensitivity
0% Load - Relieve sensor from load
or
Sensitivity - Enter the value fort the sensitivity in V
Press Key
or
and choose between 100% Load or Sensitivity
100% Load - Calibrate with 100% Load
or
Sensitivity - Enter the value for the sensitivity in mV/V
SAFE - Accept calibration values.