TecnoBody S.r.l –PROKIN Systems
Table 4 according to IEC 60601-1-2:
39
MF-PK-MANOP-02
-EN
PROKIN-MF /PROKIN-E SYSTEM – OPERATOR’S MANUAL
Guidance and Manufacturer's declaration - Electromagnetic Immunity
The Tecnobody - PROKIN SYSTEM is intended for use in the electromagnetic environment specified
below. The customer or the user of the Tecnobody - PROKIN should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
10V AM 80% 1 kHz
0.15 – 80MHz
20 V/m
80 – 1000MHz 1.4-
2GHz 2-2.5GHz
10V AM 80% 1 kHz
0.15 – 80MHz
20 V/m
80 – 1000MHz 1.4-
2GHz 2-2.5GHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Tecnobody - PROKIN, including
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d
= 1,16 √
P
d
= 1,16 √
P
80 MHz to 800 MHz
d
= 2,33 √
P
800 MHz to 2,5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol: