TecnoBody S.r.l –PROKIN Systems
Table 2 according to IEC 60601-1-2:
Note:
Ut is the a.c. main voltage prior to application of the test level.
38
MF-PK-MANOP-02
-EN
PROKIN-MF /PROKIN-E SYSTEM – OPERATOR’S MANUAL
Manufacturer's declaration - Electromagnetic Immunity
The Tecnobody - PROKIN SYSTEM is intended for use in the electromagnetic environment specified
below. The customer or the user of the Tecnobody - PROKIN should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for input
power lines
± 2 kV for
input/output lines
± 2 kV for input
power lines
± 2 kV for
input/output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line to line
± 2 kV line to earth
± 1 kV line to line
± 2 kV line to earth
Main power quality should be that
of a typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
70% 500ms
40% 100ms >95%
10ms, 5 s
70% 500ms
40% 100ms >95%
10ms, 5 s
If the user of the Tecnobody -
PROKIN requires continued
operation during power main
interruptions, it is recommended
that the Tecnobody - PROKIN be
powered from an uninterruptible
power supply.
Power frequency
magnetic field
IEC 61000-4-8
3 A/m
50 Hz
3 A/m
50 Hz
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.