10
TS580D-MM
Micro-Meter Mix Controller USER GUIDE
5.3 Air Pressure Calibration
Note: Pressure calibration must be performed when the unit is activated for
the first time.
From the Home screen button, touch the calibration button
then select
AIR PRESSURE CAL.
1. Reduce the air pressure at the rear air inlet port to zero (0) PSI. Touch the
‘LOW
(0)’ pressure measurement box. Note: Inside the yellow box is the internal
unitless pressure measurement. It will be updated.
2. Increase the air pressure at the rear air inlet port to 100 PSI (6.89 BAR). Touch the
‘HIGH (100)’ measurement box. The ‘HIGH (100)’ measurement box on the right
will be updated.
Figure 6: Calibration Select Screen
Figure 7: Air Pressure Calibration Screen