In contrast, Individual Blanking requires blanking on all wells that are to be
used for subsequent measurements. Individual blanking information is
stored for each well used and blank-correction of the samples is done with
the corresponding single blanking value (of the same well) instead of one
average blank.
Figure 6: Select Individual Blanking and click the Start Blanking button to start
individual blanking
It is recommended to use Individual Blanking as the standard blanking
option in order to obtain the most precise and reliable results.
Figure 7: Individual Blanking information in the Excel result sheet
Note
The CV value displayed in the Last Blanking box (see Figure 11
) represents
the variation coefficient of all wells used for blanking.