3. Installation Procedure
3.2 Unpacking
Procedure
1.
Visually inspect the container for damage before it is opened.
Report any damage immediately.
2.
Select a location for the instrument that is flat, level, vibration free, away
from direct sunlight, and free from dust, solvents and acid vapors. Allow 10
cm distance between the back of the instrument and the wall or any other
equipment.
3.
Place the carton in an upright position and open it.
4.
Lift the instrument out of the carton and place it in the selected location.
5.
Visually inspect the instrument for loose, bent or broken parts.
Report any damage immediately.
6.
Compare the serial number on the rear panel of the instrument with the
serial number on the packing slip.
Report any discrepancy immediately.
7.
Check the instrument accessories against the packing list.
8.
Save packing materials for future use.
3.3 Power
Requirements
The instrument is designed to operate at either 100 - 120 volt or 220 - 240 volt.
No voltage setting is required as the washer automatically senses the supplied
voltage.
The washer requires 2 x T 1.6 A / 250 V (slow blow) fuses.
WARNING
TO PREVENT THE RISK OF FIRE, THE MAINS FUSES SHOULD ONLY BE
REPLACED WITH THE SAME TYPE AND RATING OF FUSE.
3-2
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
May 2003