9.2 Power Supply Status Structure
Status and Fault Register shows the status register structure of the power supply. The Standard
Event, Status Byte, and Service Request Enable registers and the Output Queue perform standard
functions as defined in the IEEE 488.2 Standard Digital Interface for Programmable Instrumentation.
The Operation Status and Questionable Status registers implement status functions specific to
the power supply.
9.3 Condition Registers
There are two registers that the user may read to see the condition of the supply. The register
bits are set to show a fault or if an operating mode is active. The bits are cleared when the fault
or mode is cleared. The registers are read-only.
9.3.1 Fault Register
The fault register sets a bit when a Fault occurs (Refer to table 9-1). The bit is cleared when the
fault condition is removed.
Bit Number
Decimal Value
Bit Symbol
Description
0
1
1
2
AC
AC Fail
2
4
OTP
Over Temperature
3
8
FLD
Fold Back Protection
4
16
OVP
Over Voltage Protection
5
32
SO
Shut Off
6
64
OFF
Output Off
7
128
INT
Interlock
8
256
UVP
Under Voltage Protection
9
512
0
Not used
10
1024
INPO
Internal Input Overflow *
11
2048
INTO
Internal Overflow *
12
4096
ITMO
Internal Time Out *
13
8192
ICOM
Internal Comm Error *
14 to 15
N/A
0
Not used
Table 9-1: Bit Configuration of Questionable Registers
Summary of Contents for Z160-1.3
Page 5: ...This page intentionally left blank...
Page 27: ...26 Pages 26 33 are reserved for Z600 800W...
Page 28: ...27 Pages 26 33 are reserved for Z600 800W...
Page 29: ...34 2 6 Z200W 400W 600W 800W Outline Drawing...
Page 30: ...35 2 7 Z200W 400W 600W 800W Optional IEEE Isolated Analog Interface Outline Drawing...
Page 31: ...36 This Page intentionally left blank...
Page 137: ...NOTES...
Page 138: ...NOTES...
Page 139: ...NOTES...