55
[Detailed Settings] Button (For optical communication only)
Forbid Recording Start via Wireless Communication
If a Remote Unit has been set for [Forbid Recording Start via Wireless Communica-
tion], the starting / stopping of recording via wireless communication will not be
allowed for that Remote Unit. Also, makes impossible the stopping of recording via
wireless communication.
Permit Recording Start via Wireless Communication
If a Remote Unit has been set for [Forbid Recording Start via Wireless Communica-
tion], clicking here will allow the starting / stopping of recording via wireless
communication.
NOTE :
By click the [OK] button, the settings will become effective the next time the
recording is started via optical communication.
Measuring Mode (For optical communication only)
*Appears for RVR-52 only.
Voltage
(The unit of measurement on the display will show [V] for Voltage.)
Instantaneous Value --- Records the voltage at the instant of measurement.
Average Value ----------- Records the voltage as a value of the average of the
measurements within a specifi ed recording range.
Pulse
(The unit of measurement on the display will show [P] for Pulse.)
Falling -------------------Records falling pulse edge.
Rising -------------------Records rising pulse edge.
Event
(The unit of measurement on the display will show (Hi or Lo) to
signify an event and [P].)
Records the time of a pulse event.
Summary of Contents for RTR-50
Page 11: ...2 1 Introduction...
Page 16: ...7 Part Names and Functions Appearance Diagram Front Left Side Right Side Back...
Page 17: ...8 2 Getting Ready This section provides instructions on getting the RTR 50 ready to use...
Page 32: ...23...
Page 76: ...67...
Page 112: ...103...
Page 113: ...104 Others Others...
Page 120: ...111...