A : Value Adjustment Function
This function allows you correct or adjust for errors in measurement
values. The Data Logger will record only the adjusted value.
Note
- When you are carrying out adjustments and formulating equations please take care
with the settings and take personal responsibility for the process.
■
Voltage :
1
-point /
2
-point Adjustment
Before : Enter the actual measurement before adjustment
After : Enter the desired measurement after adjustment.
- Entries are valid to one decimal point.
- The input value for [Post-adjusted Reading] must be less than
±
5
V in
comparison to the input value for [Current Reading].
■
Adjustment Settings
1
.
Please follow directions as found [Communication between Data
Loggers and Computers] by connecting the RTR-
57
U unit to your
computer and place the Data Logger unit face down on top of it. (see P.
21
for detail)
2
.
In the Main Window, click on the Data Logger icon. To open the
[Adjustment] Window, click on [Settings] in the Menu Bar and then
select [Adjustment Settings] from the pull down menu.
3
.
Select either [
1
-point Adjustment] or [
2
Point Adjustment].
[Initialize] Button
[Send Settings]
Button
Select
1
-point Adjustment or
2
-point Adjustment
Enter current and
post-adjusted reading(s)
4
.
Please enter the actual pre-adjusted value(s) in the [Current Reading]
column and the post-adjusted reference value(s) in the [Post-adjusted
Reading] column.