CHAPTER 7: Updating the Firmware
7.2
Code Update Using the Library Tape
Drive Interface
The library tape drive interface (for SDLT) enables updating the policy/servo
firmware with a new version—that is to say, image—via tape update. Follow these
steps:
1.
Make certain the drive contains no cartridge.
2.
Send the CODE UPDATE REQUEST command.
3.
Send the library ATTENTION command, then check the Tape Motion Status
field of the returned General Status Packet to verify the tape drive is in the
Ready for Code Update (0x09) state.
4.
Load the cartridge containing the new firmware into the drive.
5.
Send the library ATTENTION command, then check the Tape Motion Status
field of the returned General Status Packet to verify the cartridge is loading
(0x07). Note: It takes about a minute to get into the Cup in Progress state.
6.
Send the library ATTENTION command, then check the Tape Motion Status
field of the returned General Status Packet to verify the cartridge is in the Cup
in Progress (0x0A) state.
7.
Send the library ATTENTION command, then check the Policy Firmware
Revision field to verify that the update completed successfully.
8.
Unload the cartridge and remove it from the drive.
C
AUTION
: During the firmware update, when reprogramming the
new image into the flash EEPROMs is actually in
progress, a power failure (but not bus RESET) or power
cycling the unit causes the controller module to be
unusable. When doing a firmware update, take
reasonable precautions to prevent a power failure.
Summary of Contents for SDLT220-320 INTEGRATION
Page 1: ...TANDBERG Super DLTTM Design Intergration Guide Revision 1 June 2002 432588 01...
Page 4: ...SDLT 220 and SDLT 320 Design Integration Guide...
Page 18: ...CHAPTER 1 Introduction...
Page 24: ...CHAPTER 2 General Drive Specifications Figure 2 6 Front Views of SDLT 220 320 Tape Drive...
Page 53: ...CHAPTER 4 Thermal Specifications Figure 4 3 HIM Board Thermal Measurement Location Qlogic...
Page 80: ...CHAPTER 6 SCSI and Controller Interface Specification...
Page 86: ...CHAPTER 7 Updating the Firmware...
Page 93: ......