Operation Environment

 

Place of operation 

Altitude 

Pollution Degree 

Transient  overvoltage 
at Mains Supply 

In door use only 

2000 m 

2500 V 

 

Operation Principle 

The  robot  of  SLA-E13200  uses  magnetic  rod  to  transfer 

magnetic  beads  from  one  well  to  another  and  strip  to  mix 

solution. By    moving the strip up and down to mix solution well, 

you  can  get  clean  nucleic acids  after  cells lysis,  nucleic  acids 

absorption, wash and elution.     

■ 

Flow chart for viral DNA/DNA extraction procedure 

Summary of Contents for Smart LabAssist Series

Page 1: ...User Manual Model SLA E13200 Smart LabAssist Series...

Page 2: ...ric wire directly 6 Do not place electric wires on any subject with high temperature 7 Do not look steadily at the ultraviolet lamp while it is on 8 Disconnect the plug from the power source when the...

Page 3: ...ment 07 Operation Principle 07 Intended use 08 Specifications 09 Safety and EMC information 10 Appearance of the Machine 13 Front view 13 Control panel 13 Back view 15 Internal view 15 Introduction of...

Page 4: ...3200 38 Cleaning the SLA E13200 40 Maintain the SLA E13200 42 Storage and Transport 42 Replace the fuse 42 Disposal machine 43 Chart of Pre loaded programs 44 The picture of wiring at the back of cont...

Page 5: ...www tanbead com 4 About This Manual Symbols The label on the back of the instrument the User Manual and other packaging material may contain the following symbols...

Page 6: ...ning a reagent plate and a specially designed strip the SLA E13200 can automatically handle 1 32 samples simultaneously Also the extraction process of the magnetic system is simplifier than that of th...

Page 7: ...ainless steel board The heating system is connected with the main power system Switch off both of the main power and heating switch when you turn your machine off If you are not going to start this fu...

Page 8: ...m 2 2500 V Operation Principle The robot of SLA E13200 uses magnetic rod to transfer magnetic beads from one well to another and strip to mix solution By moving the strip up and down to mix solution w...

Page 9: ...c testing Strips and reagent plates contacted with the specimen are potentially infectious After using please put them in the infectious waste bucket After extraction procedure the inside body of mach...

Page 10: ...0 Width cm Height cm Length cm Weight Kg Voltage V Current A Operation Temperature Range Operation Humidity Range 38 37 35 21 100 240 3 2 5 50 80 Temperature Range Sample Number Processing Volume Tran...

Page 11: ...quirements for Electrical Equipment for Measurement Control and Laboratory Use Part 1 General Requirements The medical device has passed the tests and conformed to the standards of IEC 61010 2 101 200...

Page 12: ...ontrol and Laboratory Use EMC Requirements Part 2 6 Particular Requirements In Vitro Diagnostic IVD Medical Equipment The medical device has passed the tests and conformed to other standards EN 61326...

Page 13: ...www tanbead com 12 Accessory Disk RS232 USB RS232...

Page 14: ...www tanbead com 13 Appearance of the Machine Front view Control Panel Introduction LED USB Plug...

Page 15: ...www tanbead com 14 Appearance of the Machine Front view Control Panel...

Page 16: ...www tanbead com 15 Back view Switches and Electrical Inlet voltage 100 240V fuse 5A 250V A Heating switch B Main power switch C RS232 port D Electrical inlet...

Page 17: ...www tanbead com 16 Internal view A Magnetic rod frame B Strip rack frame C Reagent plate rack D 8 channel strip E Reagent plate F Reagent tube A B C D E F...

Page 18: ...LED is green and others are red Start to initiate the program The buzzer will alarm when turn on the instrument Stop to terminate the program It shows STOP at status row at the first time you press t...

Page 19: ...8 Buzzer Press the button to turn buzzer beater off Pause It stops temporarily when you press the button Press the button again to continue program and the End time will be recalculated Light Turn on...

Page 20: ...een I Home Screen Touch to enter RUN Screen Touch to enter PROGRAM Screen Touch to enter CLOCK Screen Touch to enter USB Screen Touch to enter TEMP Screen Touch to enter UV LAMP Screen Touch to enter...

Page 21: ...www tanbead com 20 I Run Screen...

Page 22: ...mp The temperature is already setting and programmable It is fluctuation User could reference the reagent protocol which is written the detail Temperature Setting The temperature is setting by user cu...

Page 23: ...www tanbead com 22 II PROGRAM Screen...

Page 24: ...rogram Name The name is no more than 15 characters Program No It displays the last program number you used You can preload 100 programs program number from 00 to 99 SEARCH Insert the digits to search...

Page 25: ...www tanbead com 24 III Clock Screen...

Page 26: ...25 Year YYYY The digits from 0000 to 9999 Month MM The digits from 01 to 12 Day DD The digits from 01 to 31 Hours hh The digits from 00 to 23 Minutes mm The digits from 00 to 59 Seconds ss The digits...

Page 27: ...www tanbead com 26 IV USB Screen Import Procedure from instrument to USB...

Page 28: ...www tanbead com 27 Export Procedure from USB to instrument...

Page 29: ...www tanbead com 28 V TEMP Screen...

Page 30: ...ad com 29 Present Real working temperature Preset Set temperature of bottom stainless plate Range from RT to 70 Set Period Set up the working time hh mm Start Start time hh mm End Estimated the end ti...

Page 31: ...www tanbead com 30 VI UV LAMP Screen...

Page 32: ...www tanbead com 31 Set Period Set working time of UV Lamp hh mm Start Start time hh mm End Estimated the end time hh mm It shows current time in the right lower corner Time is 24 hour format...

Page 33: ...www tanbead com 32 Introduction of RUN Screen VOLUME The buffer volume in the well l START Start time of program hh mm ss Pause Pause time hh mm ss END End time of program hh mm ss...

Page 34: ...on with thermal couple 1 If it shows LIMIT or SENSOR it means the machine is out of order And buzzer will keep on ringing Press the button to stop the buzzer Turn the power system off and notify our s...

Page 35: ...www tanbead com 34 Introduction of execution condition 1 How to START...

Page 36: ...nger the program name will show up in the green block 2 Confirm two things before you start the program Did you put reagent plate toward outside by the missing corner and push it to the bottom of plat...

Page 37: ...nwhile other buttons will be unworkable In Collect step the robot can not be stopped immediately It will pause after finishing the collect action And the magnetic rod and strip will move back to the t...

Page 38: ...o confirm terminating the program and the robot will go back to the initial site If you don t want to terminate the program press the button to continue the program In Collect step the robot can not b...

Page 39: ...customer Pass the training of nucleic acid reagent kit 1 2 Installation Validation Before plunging into the power cable use screwdriver remove fixing plate or crew To confirm the instrument relative...

Page 40: ...rd if necessary 1 4Performance Validation Inspect the instrument in the fully state could conform the user requirements Performance Validation Contents A Consistency test Use standard reagent kit Judg...

Page 41: ...magnetic rod and aluminum rod is connecting by adhesive Please do not wiping with ethanol solution Please moisten wipes with RO water for cleaning The magnetic rod and aluminum rod is connecting by ad...

Page 42: ...ram DEMO to confirm instrument is executing in normal 1 4Regular Maintain Items List Content External inspecting i Thermostat function is normal ii Magnetic rod without broken stain iii UV Lamp functi...

Page 43: ...Operation humidity range 80 1 2 Transport To avoid damaging the linear side through the transport please use the fixing plate or fixing screws fix the mechanism before transporting Replace the fuse 1...

Page 44: ...instrument its inner region maybe will infect by specimen Wiping with 75 ethanol solution in the stainless still plate region Then turn the UV lamp sterilization 10 minutes or more to ensure that no...

Page 45: ...00 DEMO SLA E13200 01 BLOOD AUTO SLA E13200 02 VIRUS W4 AUTO SLA E13200 03 VIRUS 40 5 SLA E13200 04 B10 W4 AUTO SLA E13200 05 L BNA PK AUTO SLA E13200 06 RNA BWE AUTO SLA E13200 07 DNA RICE AUTO SLA...

Page 46: ...variable electric resistance Green 2PIN outlet J14 connecting to thermocouple P1 9PIN outlet is the usage of transmitting program P2 26PIN outlet connecting to UV lamp heater and cooler LCD contrast t...

Page 47: ...bottom of strip rack Check the plate to see if it has been pushed into the bottom of plate rack Contact service center 886 3 3607555 The action of the magnetic rod frame doesn t work normally It does...

Page 48: ...7555 Stain on the magnetic rod Be sure to put a clean strip on the strip rack whenever you operate it Avoid to be stained by the remaining reagents in the reagent tube Use clean and moisten cotton clo...

Page 49: ...ll AutoMag software 2 1 Copy AutoMag rar to your computer and extract it 2 2 If you don t have any decompression software please contact your dealer to obtain AutoMag file 2 3 You can run AutoMag soft...

Page 50: ...232 port you should connect USB connection cable to USB port Please install USB to serial converter driver If your computer is using the latest operating system you can download new USB to Serial con...

Page 51: ...www tanbead com 50 AutoMag software 1 Log in 1 1Key in user name and password 6 digits at most 1 2 Change your password 1 3 Confirm your password 1 4 Protocol Edit Window...

Page 52: ...pen an old file Save Save a file End close the file Edit Cut Copy Paste Move up Move down Insert Delete COM Port Choose a COM port There 16 COM ports COM1 is a preset port Reset a new COM port The pro...

Page 53: ...Import Import the program from a machine to a computer Excel Transfer the program data or program list to Excel Use Excel software to print the program data or program list in machine Program List Tr...

Page 54: ...ou could cut is the data in the field length that you mark 3 5 Copy Select a certain row click to copy a whole row However if you highlight a certain row you could copy all the data in the field lengt...

Page 55: ...e data in the field length 4 Introduction of program parameters Program No You could set 100 programs at most from 00 99 Use serial numbers as the names of files when saving file Program Name 15 chara...

Page 56: ...rst group is from column 1 to 6 Second group is from column 7 to 12 Temp working temperature of bottom stainless plate Input numeral from RT to 70 unit is If you input 999 it will show NA to turn off...

Page 57: ...ct ON or OFF OFF Magnetic rods stay ON Magnetic rods move down to collect magnetic beads Mixing Speed the mixing speed of strip There are four kinds of speed in the program Use them to control mixing...

Page 58: ...nd move to the top and the buzzer will ringing at the same time Stop buzzer Please press the button to stop buzz Stop pause Press the button to continue the program Vapor M The time to waiting reactio...

Page 59: ...r first then you can key in other parameters for step setting 5 3 Key in every STEP serially you couldn t skip any of them 5 4 If digits in the blue check are over the required number please press the...

Page 60: ...transmit After that click Export If it is transmitted normally a word Success will appear in the lower left corner If it is transmitted incorrectly a word Fault will appear If you see File xgp Not Fin...

Page 61: ...www tanbead com 60 Manufacturer Taiwan Advanced Nanotech Inc No 2 Aly 12 Ln 81 Longshou St Taoyuan City Taoyuan County 330 Taiwan R O C TEL 886 3 3607555 http www tanbead com...

Reviews: