![TAN Bead Smart LabAssist Series User Manual Download Page 34](http://html1.mh-extra.com/html/tan-bead/smart-labassist-series/smart-labassist-series_user-manual_3457136034.webp)
www.tanbead.com
33
:
Return to the
“Home Screen”
at the status of
END
.
:
Return to the
“Program Screen”
at the status of
END
.
SET
:
Preset working temperature.
It shows “
NA
”, when there is no preset input.
NOW
:
Real working temperature
It shows “
NC
”, when there is no connection with
thermal couple.
:
1) If it shows
LIMIT
or
SENSOR,
it means the machine is
out of order. And buzzer will keep on ringing. Press the
button
to stop the buzzer. Turn the power
system off, and notify our service center to deal with
problem.
2) If it shows
TEMP MAX
or
TEMP ALARM
, it means
heating system is out of order. And buzzer will keep on
ringing. Press the button
to stop the buzzer.
Turn heating switch off, and notify our service center to
deal with problem.
Summary of Contents for Smart LabAssist Series
Page 1: ...User Manual Model SLA E13200 Smart LabAssist Series...
Page 13: ...www tanbead com 12 Accessory Disk RS232 USB RS232...
Page 15: ...www tanbead com 14 Appearance of the Machine Front view Control Panel...
Page 21: ...www tanbead com 20 I Run Screen...
Page 23: ...www tanbead com 22 II PROGRAM Screen...
Page 25: ...www tanbead com 24 III Clock Screen...
Page 27: ...www tanbead com 26 IV USB Screen Import Procedure from instrument to USB...
Page 28: ...www tanbead com 27 Export Procedure from USB to instrument...
Page 29: ...www tanbead com 28 V TEMP Screen...
Page 31: ...www tanbead com 30 VI UV LAMP Screen...
Page 35: ...www tanbead com 34 Introduction of execution condition 1 How to START...