Installation the SLA-E13200
1.1 Installation crew:
1.1.1 Instrument installation
:
With general electronic concepts, use the
Volt-Ohm-Milliammeter for inspection.
Pass the training of Automatic Platform for Magnetic
System.
1.1.2 Reagent testing
:
With polytechnic, biomedical background, could
communicate the operation question with customer.
Pass the training of nucleic acid reagent kit.
1.2 Installation Validation
Before plunging into the power cable, use
screwdriver remove fixing plate or crew.
To confirm the instrument relative documents are
correctly and completely.
Keep a distance around 5 cm for ventilated space.
1.3 Operation Validation
To operate, inspect in normal range, then confirm
each switch and controller could not only
functioning well but also in compliance with
intended use.
Before plunging into the power cable, use
screwdriver remove fixing plate or crew.
Summary of Contents for Smart LabAssist Series
Page 1: ...User Manual Model SLA E13200 Smart LabAssist Series...
Page 13: ...www tanbead com 12 Accessory Disk RS232 USB RS232...
Page 15: ...www tanbead com 14 Appearance of the Machine Front view Control Panel...
Page 21: ...www tanbead com 20 I Run Screen...
Page 23: ...www tanbead com 22 II PROGRAM Screen...
Page 25: ...www tanbead com 24 III Clock Screen...
Page 27: ...www tanbead com 26 IV USB Screen Import Procedure from instrument to USB...
Page 28: ...www tanbead com 27 Export Procedure from USB to instrument...
Page 29: ...www tanbead com 28 V TEMP Screen...
Page 31: ...www tanbead com 30 VI UV LAMP Screen...
Page 35: ...www tanbead com 34 Introduction of execution condition 1 How to START...