ROW Precision Balance Technical Manual
- 28 -
10. ERROR CODES
Error Message
Description
Solution
-----
Maximum load exceeded Unload or reduce weight
Err 1
Incorrect date
Enter the date by using format
“yy;mm:dd”
Err 2
Incorrect time
Enter the time by using format
“hh:mm:ss”
Err 4
Zero setting error
Zero setting range exceeded
due to switching on.(4%max)
Make sure platform empty.
Err 5
Key board error
Check the keys and
connecter.
Err 6
A/D value out of range
Make sure platform empty and
check the pan is installed
proper. Check the load cell
connectors.
Err 7
Percentage Error
Please check input data, must
be > 0.5d
Err 8
Calibration weight error
Check the test weights for
calibration or linear calibration
Err 9
Unstable Reading
Check any air variation,
vibration, RF noise and
touching some where.
Check the load cell and
connecters.
Err 10
Wireless communiation
failure
Check wireless settings or
change Com settings from the
wireless
Err 11
Communication protocol
error
Check communication settings
Err 12
Accumulation error
Max accumulation times 99 /
999 / weight 999999
Err 13
Lack of unit weight
Check unit weight entry data,
must be >0.5d
Err 14
Lack of sample
Check counting samples
entry, must be >20d
Err 15
Gravity error
Check the gravity settings.
Gravity range must be 0.9xx ~
1.0xx
Err 16
Paper error
Check the printer paper
Err 17
Tare out of range
Minus weight or overload.
Remove the load and restart