AW20 Series Weighing Scale Operation Manual

- 8 -

4 Disconnect the lithium battery connector (4).
5 Replace the new lithium battery and connect the battery connector.
6 Put the foam next to the battery and close the battery cover

2

3

4

1

Summary of Contents for AW20

Page 1: ...Operation Manual AW20 Weighing Scale A1...

Page 2: ......

Page 3: ...Log in 13 5 3 Zero Setting 14 5 4 Tare 14 5 4 1 Semi automatic tare 14 5 4 2 Automatic tare 14 5 4 3 Pre set tare 15 5 5 Weighing Operation 17 5 6 Weighing Units Conversion 18 5 7 PLU Information Stor...

Page 4: ...its setting for target percentage 36 5 12 5 Tolerance Checking Mode 37 5 13 Backlight Setting 38 5 14 Auto Power Off 39 5 15 Animal Weighing 40 5 16 Peak Hold 41 6 CALIBRATION 43 6 1 Normal Calibratio...

Page 5: ......

Page 6: ...LING CLEANING OR SERVICING FAILURE TO DO SO COULD RESULT IN BODILY HARM OR DAMAGE THE UNIT CAUTION Permit only qualified persons to service the instrument Before connecting or disconnecting any compon...

Page 7: ...this device Verify the local voltage and receptacle type are correct for the scale Only use original adaptor other could cause damage to the scale Pluggable equipment must be installed near an easily...

Page 8: ...t percentage weighing tolerance checking Accumulation mode with manual accumulation auto load accumulation and auto unload accumulation Up to 300 PLUs capacity 8 direct PLUs supported to display alpha...

Page 9: ...AW20 Series Weighing Scale Operation Manual 4 2 2 Over View 1 2 3 4 5 6...

Page 10: ...ption 1 RJ45 Support 10M 100M Ethernet 2 USB2 0 For USB disk scanner data transfer don t use for charging other device 3 RS232 For connecting computer printer alarm light Remote display Second platfor...

Page 11: ...cale later Inspect the scale for damage Make sure all components are included 1 Scale 2 Adaptor 3 Manual 4 Stainless steel pan 3 2 Installation Remove transport lock from bottom of the scale base Plac...

Page 12: ...will flicker on the display Approximately 30 minutes of instrument usage are left afterwards it will shut off automatically Please use the supplied battery charger for charging the battery immediatel...

Page 13: ...ghing Scale Operation Manual 8 4 Disconnect the lithium battery connector 4 5 Replace the new lithium battery and connect the battery connector 6 Put the foam next to the battery and close the battery...

Page 14: ...Unit Weight key Used to store the known average piece weight of item Long press to change the average piece weight value of item Sampling key Used to enter the counting operation Product key Used to r...

Page 15: ...ighing mode Tare key Used to perform a tare function Subtracts weights Limit key Used to enter check weighing mode Long press to set HI and Low limits Accumulation key Used to store the accumulation w...

Page 16: ...ght Indicates scale is displaying the net weight Tare weight Indicates scale is displaying the tare weight Gross weight Indicates scale is displaying the gross weight Lack of unit weight Indicates pie...

Page 17: ...AW20 Series Weighing Scale Operation Manual 12 Total Displays accumulated total weight Range Display Bar Displays the range of capacity...

Page 18: ...s on it may be asked to input the account and password to log in 2 Input the account by using numeric keys press key to confirm 3 Input the password by using numeric keys press key to confirm 4 Displa...

Page 19: ...It will be shown with a minus value 4 This Tare value can be cleared by pressing key when there is no load on the platter This tare entry cannot take place if the platter is in in motion 5 4 2 Automa...

Page 20: ...are window will be active 2 Using numeric keys to input the tare value 3 Press key to confirm the preset tare value appears Indirect pre set tare The scale has a pre set tare function it can be combin...

Page 21: ...ct PLU list 2 Display will be show selected product name number pre tare value 3 Place the item on the pan and wait to get weight stable display show net weight 4 Remove item from the pan the net wind...

Page 22: ...4 Remove item from the pan the net window show negative value 5 Press key to clear tare value 5 5 Weighing Operation Switch on the scale display show normal weighing mode Load the weight on the pan di...

Page 23: ...lative units symbol will turn on in the display To enable the weighing units please see parameter 2 ADVANCED SETTINGS 1 DISPLAY UNIT 1 Press key to enter setup menu 2 Press key to select 2 ADVANCED SE...

Page 24: ...r to setup menu 2 PRODUCT INFO 2 ADD EDIT PRODUCT Operation 1 Press key to enter setup menu 2 Press key 2 times to select 2 PRODUCT INFO 3 Press key to enter 4 Press key to select 2 ADD EDIT PRODUCT 5...

Page 25: ...ng the current product number name or by selecting a new PLU key will be clear existing PLU number name automatically 5 7 3 Store Direct PLU information 1 Press key to recall the indirect PLU informat...

Page 26: ...AW20 Series Weighing Scale Operation Manual 21 5 7 4 Recall direct PLU In normal weighing mode press M key selected from to to recall direct PLU...

Page 27: ...tion scale should be stable and return to zero sampling available only when weight more than 20d 5 8 1 Manual Sampling for counting procedure 5 8 1 1 Simple Sampling Make sure platter empty and displa...

Page 28: ...r weight and number of items will be increased Remove the items the average piece weight will remain stored in memory for the next counting of similar pieces Press and hold key to back to normal weigh...

Page 29: ...ample window will be active Using numeric keys to input the count number eg 10 Press key to confirm display show dotted line The average piece weight of item will be calculated and then the number of...

Page 30: ...ribed Press and hold key to back to normal weighing mode Note The weighing may be amended on the basis of greater numbers of samples improves the accuracy of counting large quantities 5 8 2 To input t...

Page 31: ...ly add further parts whose number is smaller than that of the first reference determination For each reference optimization the average piece weight is newly calculated As the additional pieces increa...

Page 32: ...rcentage of 0 47kg 100 Place 0 210kg weight on the platform display will be show 44 7 as per the percentage of 0 470kg 100 Press and hold key to back to normal weighing mode The weighing may be amende...

Page 33: ...mulation available only when weight more than 20d 5 10 1 Manual Accumulation 1 Place the load on the platform the weight display will be show the net weight 2 Press key when STABLE indicator on 3 Tota...

Page 34: ...lay get zero and stable then place the second weight 5 It can continue until the memory gets fully 5 10 3 Auto Unload Accumulation To enable the auto unload accumulation please refer to parameters 2 A...

Page 35: ...ation records will be show 3 Press key or key to scroll up or scroll down to view the records 4 If connected with printer press key to print out the accumulation records 5 Press key to back to normal...

Page 36: ...tion and No Printing If connected with printer the data will be print out in accumulation mode see parameters 2 ADVANCED SETTINGS 7 ACCUMULATION 3 PRINT EVERY RECORD NO MANUAL Load the weight after st...

Page 37: ...s NOTE When switching between products with counting modes there will be warning window to clear the total The user has to clear the total by pressing long CE from the ACC report 5 11 Printing 5 11 1...

Page 38: ...ighing goods according to a target quantity or target weight or target percentage within specified tolerances With this function one can also check if the weighing good is within a defined tolerance r...

Page 39: ...lerance range Target quantity target weight target percentage exceeds maximum tolerance limit 5 12 2 Limits setting for target weight 1 Press and hold key to enter limits setting in Normal weighing mo...

Page 40: ...ion enter zero in both limits 5 12 3 Limits setting for target quantity 1 To enter counting mode by sampling or inputting the known unit weight of item 2 Press and hold key to enter limits setting 3 T...

Page 41: ...imits 5 12 4 Limits setting for target percentage 1 Press key to enter percentage weighing in normal weighing mode 2 Press key to enter limits setting 3 The lower limit value in the display will be ac...

Page 42: ...ight more than 20d The function is not valid when one enters a lower limit greater than the upper limit When lower limits are the same with upper limits display only shows one set value 5 12 5 Toleran...

Page 43: ...ter to system settings 3 Press key until to select the backlight 4 Press key to enter the menu 5 Press key to change the setting LOW Backlight turn to be low light NORMAL Backlight turn to be medium l...

Page 44: ...riod of time in minutes Operation 1 Press key to enter setup menu 2 Press key to enter to system settings 3 Press key until to select the power sleep 4 Press key to enter the menu 5 Press key to chang...

Page 45: ...5 Press key to enter settings 6 Press key to change the settings until display show ANIMAL WEIGH 7 Once selected the animal weighing mode it could enter 5 ANIMAL WEIGH sub menu to set the relative ran...

Page 46: ...ears with a beep sound the final weight will be shown on the display The weight will be update when the load weight exceed the floating range and jump range To disable the animal weighing function it...

Page 47: ...settings 6 Press key to change the settings until display show PEAK HOLD 7 Press key to confirm 8 Press key two times to exit the setup menu and back to peak hold weighing mode 9 The Peak indicator wi...

Page 48: ...rameters 4 SCALE1 2 Calibration 1 Normal Calibration Before access to the parameter it need to press calibration switch locating back side of indicator in approval version Steps Remark 1 Press Key to...

Page 49: ...the platter Better to use minimum 1 3 of the scale capacity 6 Press key to confirm The display will be show Please empty the scale 7 Remove the test weight from the scale 8 Press key to confirm The di...

Page 50: ...g and Stainless Steel pan Interface RS232 RJ45 USB2 0 Optional RS485 WIFI Bluetooth analog output second platform Operation Temperature 10 C 40 C 14 F 104 F Power AC Adaptor 12V 2500mA Optional Rechar...

Page 51: ...inless Steel pan Interface RS232 RJ45 USB2 0 Optional RS485 WIFI Bluetooth analog output second platform Operation Temperature 10 C 40 C 14 F 104 F Power AC Adaptor 12V 2500mA Optional Rechargeable Li...

Page 52: ...ck the load cell connectors Sample low Percentage Error Please check input data must be 1 of full capacity Lack of unit weight Check unit weight entry data must be 0 1d Lack of sample Check counting s...

Page 53: ...device 4 Do not use organic acid alkaline solutions scouring powders or plastic dissolving cleaning agents for cleaning 5 Do not spray directly on the device 6 Avoid strong pressure or scratching the...

Page 54: ...ibration Check pan is installed correctly Verify installed in an acceptable location Cannot use Full Capacity Overload Stopper touching pan support or bottom of the load cell Transportation lock not r...

Page 55: ......

Page 56: ...TScale Electronics Mfg Kunshan Co Ltd 99 Jingwei Road Kunshan Jiangsu China Tel 86 512 57669080 www t scale com...