AW20 Series Weighing Scale Operation Manual
- 47 -
8. ERROR CODES
Error Message
Description
Solution
ERROR 4
Zero setting error
Zero setting range exceeded due to
switching on.(4%max)
Make sure platform empty.
ERROR 5
Key board error
Check the keys and connector.
ERROR 6
A/D value out of range
Make sure platform empty and check
the pan is installed proper. Check the
load cell connectors.
Sample low
Percentage Error
Please check input data, must be >
1% of full capacity
Lack of unit weight
Check unit weight entry data, must
be >0.1d
Lack of sample
Check counting samples entry, must
be >20d
ERROR 19
Initialize
zero error
Calibration the scale.
ERROR 20
Wireless communication
failure
Check wireless settings or change
Com settings from the wireless
--UL--
Under load
Minus weight, check the platform and
restart or calibrate.
--OL--
Over range
Remove the load.
Re calibrate
BAT LOW
Low battery indicator
Please charge the battery