Fix transverse connector
First tighten the fixing screws for the rod of both cross-link
clamps firmly with the hexagonal screwdriver
4.0 mm with
T-handle (1). Then tighten both set screws of the
3.5 mm
cross-link rod firmly with the hexagonal screwdriver
2.5 mm (2).
4
Distract cross-link assembly (optional)
Loosen one of the set screws with the small hexagonal
screw driver, place the holding forceps next to the relevant
clamp and carry out the distraction with the spreader forceps.
Retighten the set screws.