Synthes
7
5
Remove the handpieces and the attachments
To remove an Electric Pen Drive handpiece, turn the release
sleeve as indicated by the arrow and remove the handpiece.
To remove the Air Pen Drive handpiece, simply turn the
handpiece anticlockwise and remove from the Adapter.
When the Maintenance Unit is only used for maintaining the
Air Pen Drive it is possible to keep the Adapter in the Mainte-
nance Unit.
The attachments can be removed by turning them clockwise
in the connector and carefully pulling them out.
After the sterilizing process the handpiece and the attach-
ments will be ready to use.
4
Start the lubrication process
Before starting the lubrication process the Lid must be
closed. After closing the Lid, push the START button to start
the lubrication procedure. The lubrication procedure will take
approximately three minutes when all connectors are filled.
The READY LED will light when the lubrication loop is
successfully completed.
If the ERROR
LED flashes or lights up, consult the chapter
Troubleshooting.
Note:
Make sure the oil is filled up.