8
Synthes
Maintenance Unit. For EPD and APD
Instructions for Use
Care and Maintenance
2
Change the filter
If oil can be seen in the filter in the inspection window, the
filter should be changed. We recommend changing the filter
every time the oil has to be filled up.
To change the filter, open the Lid, remove the old filter and
fit the new filter into the bottom of the Maintenance Unit.
The oily filter and the empty bottle of Synthes special oil,
for EPD and APD, should be disposed according to national
regulations.
Note:
The Maintenance Kit 05.001.094 contains 1 bottle of
Synthes Special Oil 05.001.095 and 1 spare filter.
1
Fill the oil tank
If the oil tank is empty, the ERROR
LED will stay lit up; oil
should then be refilled. Open the oil tank by turning the
cap counterclockwise and refill only with Synthes Special Oil,
50 ml, for EPD and APD, 05.001.095.
Note:
Only use the provided Synthes Special Oil, for EPD and
APD, 05.001.095. Other oils can have a toxic effect and can
lead to sticking. In this case warranty claims will be invali-
dated.