3
The Lubrication Process
1
Installation
Install the provided Power Supply in the power supply con-
nection, then connect the country specific Power Cord to the
Power Supply and to a plug socket.
First, applying slight pressure, plug the simple end of the
Pneumatic Connection into the air connector on the back of
the module; then connect the nipple for the Pneumatic
Connection to the air supply.
Fill the oil tank with Synthes special oil 50 ml for EPD and
APD (05.001.095), as described in the chapter Care and
Maintenance.
Note:
To completly disconnect from AC mains disconnect
the power supply.
2
Start the Maintenance Unit
Push the ON/OFF button
on the keypad to turn on the
Maintenance Unit. The POWER ON LED should stay lit up.
If the ERROR
LED flashes or lights up, consult the chapter
Troubleshooting.