Synthes
5
Connect the Air Pen Drive handpiece (05.001.080)
To connect the Air Pen Drive handpiece, the Adapter for
Maintenance Unit for Air Pen Drive (05.001.089) has to be
used. Place the female side of the Adapter on the connector
on the left of the Maintenance Unit, making sure that the
pins on the connector engage in the slotted link on the
Adapter and then turn the Adapter clockwise. To apply the
handpiece, set the adjustment sleeve to the foot switch posi-
tion. Then fit the air hose side of the handpiece to the con-
nector by fitting the pins on the Adapter into the grooves in
the air hose coupling side of the handpiece and turn the
handpiece clockwise.
Caution
– The Air Pen Drive must be set to foot switch position
during the lubrication process!
– The Air Pen Drive has to be lubricated through the air
inlet/outlet, not through the attachment coupling!