CORRECTION MANEUVERS
SEGMENTAL TRANSLATION
Equipment
2997-04-175
Rod Clamp
2997-04-180
Double End 5.5 Rod Hex Wrench
2997-04-170
French Rod Bender
2997-04-220
Key Inserter/Tightener
2997-04-230
X25 Inserter/Tightener
2997-04-145
Pencil Handle
2997-04-205
Facilitator
Alternative Equipment
2997-04-185/190
Coronal In-Situ Bender, Left/Right
2997-04-195/200
Sagittal In-Situ Bender, Left/Right
2997-04-355
Tab Ring
Segmental Translation is a widely used maneuver for the
correction of coronal deformity. The main goal of this
technique is to segmentally bring the spine to the contoured
rod, allowing for coronal correction to occur.
1,2,11
The EXPEDIUM VERSE Spinal System provides extended
tabs with 17mm of reduction. The combination of the
reduction tabs and the hypermobility or extended range of
polyaxial motion simplify rod capture. Consequently, the
use of the EXPEDIUM VERSE Screw at each level permits a
more uniform distribution of the forces required for
correction, reducing the magnitude of stresses at the
bone-screw interface and the potential of screw
migration.
1
This maneuver can be performed in
conjunction with other techniques to allow for sagittal,
coronal, and axial correction.
(FIGURE 33-35, 37-39)
FIGURE 33
FIGURE 34
FIGURE 35