Unitized Set Screw
One-step locking of the polyaxial head
and rod can be achieved with the
Unitized Set screw when independent
locking is not required.
Correction Key
The Correction Key delivers independent
locking of the polyaxial head and rod,
providing the benefit of
a polyaxial screw with the correction
control of a monoaxial screw.
Rod Lock
Poly Lock
EXPEDIUM VERSE Polyaxial
Pedicle Screw
TOP NOTCH
®
Feature
The Top Notch Feature helps
various instruments easily connect
to the implant and simplifies
intraoperative maneuvers.
Head Drag
A drag mechanism enables the
implant head to retain alignment,
simplifying rod capture
Hypermobile Head with a
Favored Angle
The favored angle provides up to 84°
of angulation,* providing the benefits of
a side loading implant with the ease
of use of a top tightening system.
X25 Drive Feature
For the first time in an EXPEDIUM 5.5
Pedicle Screw, the EXPEDIUM VERSE
System provides an X25 Drive Feature
which allows for 43%** more torque
to be delivered compared to the
T20 Drive Feature provided in the
EXPEDIUM 5.5 System.
Fully Threaded Dual-Lead Shank
The double-lead thread, self-tapping and
self-centering shank is designed to
provide faster screw insertion, enhanced
bone purchase, and help ensure
accurate placement within the pedicle.
FIGURE 2
FIGURE 3
Extended Tabs with Loading Zone
Powerful integrated reduction with an
alignment feature.
The black line on the tab indicates the side
of the Favored Angle.
* DVA-106873-TA
** DVA-102149-TP Rev 4
DePuy Synthes Companies
|
EXPEDIUM VERSE System
|
System Guide
2
Summary of Contents for EXPEDIUM VERSE
Page 1: ...SYSTEM GUIDE EXPEDIUM VERSE Spinal System...