NOTE:
It is essential to confirm that cement flow
has stopped before disengaging the cement delivery
system from the screw head.
CEMENT DELIVERY
Follow the instructions from the respective bone cement and
delivery system package inserts to introduce the cement
through the delivery cannula.
Use fluoroscopy throughout the procedure to verify and
monitor cement flow as appropriate.
REMOVAL OF DELIVERY CANNULA
When the appropriate amount of cement has been
introduced, stop cement introduction as indicated per the
respective bone cement technique.
Disengage the cannula from the alignment guide by
depressing the tabs on each side of the cannula and remove
it from the screw as soon as cement injection is completed
and flow has stopped through the cannula.
FIGURE 23
FIGURE 24
NOTE:
Controlled delivery is essential to proper
screw augmentation. Overly aggressive cement
injection may result in cement leakage and
unsatisfactory results. Immediately stop cement
injection if extravasation is detected.
15
DePuy Synthes Companies
|
EXPEDIUM VERSE System
|
System Guide