Safety Information
NuGenius/
N
User manual
10
-
Has been subjected to severe stress during transportation
Electrical protection
Insulation: Class I rating for external circuits. Only connect equipment that
meets the requirements of IEC 61010-1, IEC 60950 or equivalent standards.
Installation Category: The instruments are able to withstand transient over
voltages typically present on the MAINS supply. The normal level of transient
over voltages is impulse withstand (overvoltage) category II of IEC 60364-4-
443.
Pollution Degree 2: Normally only non-conductive pollution occurs.
Occasionally, however, temporary conductivity caused by condensation
must be expected.
Electrical Specifications
The NuGenius/N power rating is as follows:
Voltage
100-240Vac
Frequency
50-60Hz
Current
2A
Fuse
2 of 20 mm x 5 mm IEC127 T 6.3 A H 250V
Changing fuses
There are two sets of fuses that you may need to replace: the first set is located
near the electrical power cord socket on the rear of the darkroom; the second
set is located on the rear of the UV Transilluminator inside the darkroom (if
supplied).
To change the fuses on the rear of the darkroom:
Switch off the darkroom and remove the line power cord from the electrical
supply.
Gently pull out the fuse holder on the rear of the darkroom
:
Replace the fuse(s) with new ones of the same type and rating. The fuse
type is 20 mm x 5 mm IEC127 T 6.3 A H 250V for instruments in all
countries.
Replace the fuse holder.
Power socket
Fuse holder