Page 15
CALIBRATION
•
This parameter allows the user to calibrate the balance any
time deemed necessary.
•
Switch the balance off and then on again by pressing
[On/Off]
. During the self-test press
[PCS]
and
[Print]
together.
•
Display will show "
unLoAd
". Remove all weights from the
pan.
•
Wait for the stability indicator to be on and then press
[Zero].
•
The display will then show the first calibration weight request.
•
Press
[Unit]
to scroll through the other calibration weights.
•
Press
[Zero]
to select calibration weight. The display will
show “
LoAd
”.
•
Place this weight on the platform and press
[Zero]
when the
stability indicator will be on to show the value is stable.
•
If the calibration is complete, the balance will show “
PASS
”
and will show the weight put on the balance.
•
If the wrong calibration weight is used, the balance will show
“
FAlL H
” (when the weight is higher) or “
FAlL L
” (when the
weight is lower). Repeat the process using the correct
calibration weight.
Calibration weights:
Model #
120g
400g
800g
2000g
4000g
Weight
1
60g 200g 400g 1000g 2000g
Weight
2
120g 400g 800g 2000g 4000g
Weight
3
COMMUNICATION WITH A COMPUTER/PRINTER
These Compact Portable Balances come with a standard RS-232
interface.
The Standard Interface Parameters are:
Connection Details are:
Connector: 9 pin d-subminiature female
socket
Pin 3 Output from the balance
Pin 2 Input to the balance
Pin 5 Signal Ground
RS-232 output of weighing data
ASCII code
4800 Baud
8 data bits
No Parity