5120A/5115A Operations and Maintenance Manual
39
Figure 12: The Test Mask Editor
Adding or editing points requires inputting x and y coordinates. The number pad editor is executed
twice each time a point is added or edited. When adding new points, the number pad editor is
initialized with blank fields. When editing existing points, it is initialized with each coordinate’s
current value. This simplifies modification of only one of an existing point’s coordinates.
To edit a test mask:
1. Access the test mask editor as described in
“ To access the test mask editor:”
2. Press
Edit Mask
.
The softkey interface changes to provide an interface with the interactive table.
3. Press
Add Point
to add a new point to the test mask or go to
to edit existing points
or
to delete points.
Tip: Right-click the test mask table and select
Add Point
in the menu without entering
the
Edit Mask
interface.
“3.4.1.1 The Number Pad Editor”
on page 17 is displayed for the first coordinate.
•
Enter the
Offset Frequency
coordinate (in Hz) for the new point and press
Apply
,
or press
Cancel
to cancel point addition.
If the first coordinate was applied, the number pad editor is displayed for the second
coordinate.
•
Enter the
dBc/Hz
coordinate for the new point and press
Apply
, or press
Cancel
Current mask
is highlighted
Toggles visibility of the current
test mask. Displays
Show Mask
when mask is hidden, or
Hide
Mask
when mask is visible.
The test mask editor provides the softkey
interface to edit, load, and save test masks.
Press
Edit Mask
to edit the
current test mask.
Summary of Contents for 5115A
Page 6: ...iv Contents Index 97 ...
Page 12: ...6 1 Introduction ...
Page 16: ...10 2 Setting Up a PNTS ...
Page 26: ...20 3 The Screen and Basic Concepts ...
Page 82: ...76 9 Maintaining a PNTS ...
Page 92: ...86 Appendix A Specifications ...
Page 98: ...92 Appendix B Theory of Operation ...
Page 100: ...94 ...
Page 102: ...96 ...
Page 106: ...100 ...