26
A-96.250.351 / 230320
AMU Rescon
Instrument Setup
USP Operating
Mode
Menu 5.1.3
The USP Operating Mode implemented in the firmware of the AMU
Rescon transmitter allows the measurement of pharmaceutical wa
-
ter according to the USP <645>.
If the USP Operating Mode is set to <off> conductivity or resistivity
standard measurement is performed.
If the USP Operating Mode is set to <on>, the uncompensated
measuring values are compared with the values of an implemented
table defined by USP. If the deviation is too high, the Error 15 (USP
Error) is issued.
Sensor
Parameters
Menu 5.1.4:
Enter the following parameters printed on the sensor label.
Cell constant ZK
Temperature correction DT
Sensor cable length. If the sensor cable length is 0.3 m, set the
cable length to 0 m.
Temp.
Compensation
Menu 5.1.5
Choose between:
none
coefficient
neutral salts
high purity water
strong acids
strong bases
ammonia, ethanolamine
morpholine
87.322.210
SW2540001 ZK = 0.00997
DT = -0.13 °C
RC U, 0.3 m