14
A-96.250.351 / 230320
AMU Rescon
Installation
3.
Installation
3.1. Installation Checklist
Check
Instrument’s specification must conform to your power ratings.
Do not switch on power until all external devices are connected.
Installation
The transmitter is intended for panel mounting. The dimensions
Dimensions of the AMU Transmitter, p. 15
Electrical Con
-
nections
Connect all external devices, see
.
Connect the power cord, but do not switch on power until all ex
-
ternal devices are connected.
Connect sen
-
sors
.
Power-up
.
Switch on power.
First the instrument performs a self-test, displays the firmware
version and then starts normal operation.
Instrument
set-up
Program all necessary parameters see
Program the flow cell type
Program the measuring mode (resistivity or conductivity).
Program all sensor data (cell constant, temperature correction,
cable length).
Program the required temperature compensation.
Program all parameters for external devices (interface,
recorders, etc.).
Program all parameters for instrument operation (limits,
alarms).
Run-in period
If the conductivity value of the sample is very low, the sensor
might need some time until the correct reading is displayed