AMU Rescon
Default Values
10. Default Values
NOTICE:
The AMU Rescon has two different measuring modes
(Resistivity or Conductivity) which can be set in menu
<Installation>/<Sensors>/<Meas. Mode>. The instrument
remains in the selected operating mode even after the <Default
Values> are reset completely. Therefore, this default value list is
divided in the two parts Resistivity and Conductivity where
necessary.
Operation:
Sensors:
Filter Time Const.:...................................................................... 20 s
Hold after Cal.:......................................................................... 300 s
Alarm Relay
...................................................................... same as in Installation
Relay 1 and 2
...................................................................... same as in Installation
Input
...................................................................... same as in Installation
Logger:
Logger Interval:..................................................................... 30 min
Clear Logger:................................................................................no
Installation:
Sensors
Flow: ........................................................................................ None
Meas. Mode:.................. remains as set, Resistivity or Conductivity
USP Operating Mode:.................................................................. Off
Sensor parameters
Cell Constant:.......................................................... 0.01000 cm
-1
Temp. Corr.........................................................................0.00 °C
Cable length .........................................................................0.0 m
Temp. Compensation
Comp.................................................................................... None
Signal Output 1
Parameter: .................................................................... Meas. Value
Current loop:.....................................................................4 –20 mA
Function:.................................................................................. linear
Resistivity
Scaling: Range low:............................................................ 0.00 MΩ
Scaling: Range high:......................................................... 20.00 MΩ
Conductivity
Scaling: Range low:........................................................... 0.000
µ
S
Scaling: Range high:........................................................... 1000
µ
S
Signal Output 2
Parameter: ....................................................................Temperature
Current loop:.....................................................................4 –20 mA
Function:.................................................................................. linear