AMU Rescon
Instrument Setup
A-96.250.351 / 230320
25
4.
Instrument Setup
4.1. Establish Sample Flow
1
Open the flow regulating valve.
2
Wait until the flow cell is completely filled.
3
Switch on power.
NOTICE:
To allow a precise measurement in the range
between 18–18.18 MΩ, set the sample flow to 70–100 l/h.
4.2. Programming
Set all necessary parameters in menu 5 <Installation>, further infor
-
mation about sensor parameters see
Flow measurement
Measuring Mode
USP Operating Mode
Sensor parameters
Temperature compensation
Program all parameters for external devices (interface, re
-
corders, etc.). Program all parameters for instrument opera
-
, for
explanations, see
Program List and Explanations, p. 47
Flow
Measurement
Menu 5.1.1
Select the flow sensor you are using:
None
Q-Flow
Q-HFlow
Q-Flow can be used for flow rates up to 25 l/h.
Q-HFlow can be used for flow rates up to 120 l/h.
Meas. Mode
Menu 5.1.2
Select Resistivity or Conductivity according to your application.