AMI Turbitrace
Program List and Explanations
5.4.3
Load Firmware:
Firmware updates should be done by instructed
service personnel only.
5.4.4
Password:
Select a password different from 0000 to prevent unau
-
thorized access to the following menus:
5.4.4.1
Messages
5.4.4.2
Maintenance
5.4.4.3
Operation
5.4.4.4
Installation.
Each menu may be protected by a
different
password.
If you forgot the passwords, contact the closest SWAN representative.
5.4.5
Sample ID:
Identify the process value with any meaning full text,
such as KKS number.
5.4.6
Line Break Detection:
Define if message E028 should be issued in
case of a line break on signal output 1 or 2.
Choose between <Yes> or <No>.
5.5 Interface
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
Load Firmware
no
yes
5.5.1
Protocol:
Profibus
5.5.20
Device address:
Range: 0–126
5.5.30
ID-Nr.:
Range: Analyzer; Manufacturer; Multivariable
5.5.40
Local operation:
Range: Enabled, Disabled
5.5.1
Protocol:
Modbus RTU
5.5.21
Device address:
Range: 0–126
5.5.31
Baud Rate:
Range: 1200–115 200 Baud
5.5.41
Parity:
Range: none, even, odd
5.5.1
Protocol:
USB-Stick
Only visible if an USB interface is installed. No further settings are
possible.
5.5.1
Protocol:
HART
Device address:
Range: 0–63