44
A-96.250.131 / 050220
AMI Turbitrace
Maintenance
Calibration
flow
Manual
calibration
A calibration can be started manually or automatically.To start a
manual calibration proceed as follows:
Automatic
calibration
To program an automatic calibration proceed as follows:
LINE
CAL
Line
Zero
Empty
Cal
EMPTY
A
B
1
Navigate to menu <Maintenance>/
<Calibration>.
2
Press [Enter].
3
Select <Parameters>.
4
Press [Enter].
5
Select <Method>.
If measuring turbidity in pure
and ultrapure water, set method
to <fine>. For measuring
turbidity >0.5 FNU set method
to <coarse>.
6
Press [Enter], then press [Exit
7
Select <Yes> to confirm.
8
Press [Enter]
9
Select <Manual operation>
10
Press [Enter]
The calibration starts.
3.2
Verification
Maintenance
Set Time 01.01.05 16:30:00
Simulation
Calibration
3.2.2
Calibration
Parameters
Manual operation
3.2.2.2
Mode
daily
Start Time
00:30:00
Calibration
Run Time
300 sec
Method
fine
coarse
Method
fine
1
Navigate to menu <Maintenance>/
<Calibration>.
2
Press [Enter].
3.2
Verification
Maintenance
Set Time 01.01.05 16:30:00
Simulation
Calibration