38
A-96.250.131 / 050220
AMI Turbitrace
Maintenance
Replacement
procedure
To replace the humidity absorber proceed as follows:
1
Select <Maintenance>/<Replace Absorber> and confirm with
[Enter].
This makes the two messages E068 and E006 disappear for
three days.
2
Unscrew and remove the screw cover [A] from the measuring
cell [E].
3
Pull the flask [C] containing the saturated humidity absorber out
of the screw cover.
4
Remove the cover of the new humidity absorber flask [B].
5
Push the new humidity absorber flask into the screw cover.
A
B
C
Screw cover
New humidity absorber flask
Saturated humidity absorber
D
E
Bayonet lock cover with
o
-
ring
Turbidimeter housing
A
B
D
E
C