AMI Turbitrace
Maintenance
A-96.250.131 / 050220
45
3
Select <Parameters>.
4
Press [Enter].
5
Select <Mode> and press [Enter].
6
Select the <Mode> according your
requirements.
Additionally, the Modes <Input>
and <Fieldbus> are available
which you can chose by
scrolling down with the
[
]
key.
7
Select <Yes> to confirm.
8
Select <Method>.
If measuring turbidity in pure
and ultrapure water, set method
to <fine>. For measuring
turbidity > 0.5 FNU set method
to <coarse>.
9
Select <Run time> and set it ac
-
cording your requirements.
10
Depending on your settings set the
interval or the start time.
3.2.2
Calibration
Parameters
Manual operation
3.2.2.2
Method
fine
Start Time
00:30:00
Calibration
Run Time
300 sec
Mode
daily
weekly
Mode
daily
interval
off
3.2.2.2
Mode
daily
Start Time
00:30:00
Calibration
Run Time
300 sec
Method
fine
coarse
Method
fine
3.2.2.2
Mode
daily
Start Time
00:30:00
Calibration
Method
fine
Run Time
300 sec