AMI Turbitrace
Maintenance
A-96.250.131 / 050220
35
6.
Maintenance
6.1. Maintenance Schedule
Preventive maintenance frequency depends on water quality, on
the application, and on national regulations.
Every week
Check sample flow.
If necessary
Clean the measuring chamber.
Replace the 0-point filter
When maintenance
message E068
appears
Replace the humidity absorber.
Recommendation:
When replacing the humidity absorber,
check also the O-ring of the bayonet lock
cover and replace it if necessary.