6 - Maintenance - Repair
Replacing Parts
6.4.7 Flow
sensor
1
Shut down the instrument according to
6.3.1,
36
.
2
Open the transmitter cover (4 captive screws).
3
Disconnect flow sensor cable from terminal. Cable
designation see
8.3.2
“Connections Diagram”,
74
.
4
Loosen cable gland nut and retrieve cable.
5
Loosen the two
screws fixing the flow
cell to the base plate.
6
Pull the flow cell
slightly off the base
plate and unscrew
sensor from
measuring cell.
7
Install in reverse
order. Tighten sensor
fingertight only.
AMI Trides Operator’s Manual, A-96.250.101
47