![Swan Analytical Instruments A-85.119.010 Manual Download Page 113](http://html1.mh-extra.com/html/swan-analytical-instruments/a-85-119-010/a-85-119-010_manual_1411374113.webp)
8 - Installation
Instrument Setup
Timer
Typical use: Flow cell cleaning.
Example: To flush the system with cleaning solution for 5 min
every hour.
Mode
interval
5.3.2.24
Mode
daily
5.3.2.24
Mode
weekly
5.3.2.24
Run Time
5.3.2.44
1
– 1440 min, default 1 min
Interval for the Run Time to occur.
00:00:00
– 23:59:59 min, default 00:00:00
Daily time when Run Time is to occur.
Start time 00:00:00 - 23:59:59 min,
default 00:00:00
Monday - Sunday on/off, default all off
Time and day(s) of the week when Run Time is to
occur.
5
– 6000 sec, default 10 sec
The length of time Run Time is active (i.e. the
relay is closed).
Delay
5.3.2.54
0
– 6000 sec, default 5 sec
Time by which the execution of Run Time is
retarded after receiving the signal to start.
Signal Outputs
5.3.2.6
continuous
- Signal Outputs are active at all
times (default).
hold
- During Run Time, Signal Outputs are on
hold, i.e. the value indicated just before the
start of Run Time is retained.
Output/Control
5.3.2.7
off
- No signal on Signal Outputs (set to 0).
continuos
- Control is active at all times (default).
hold
- During Run Time, values determining con-
trol are on hold, i.e. dosing is performed according
to these frozen values.
off
- Control deactivated.
AMI Trides Operator’s Manual, A-96.250.101
105
Summary of Contents for A-85.119.010
Page 10: ...1 AMI Trides keys 2 AMI Trides Operator s Manual A 96 250 101...
Page 15: ...2 Safety Instructions General 2 2 1 AMI Trides Operator s Manual A 96 250 101 7...
Page 20: ...3 3 1 2 AMI Trides pH 3 2 3 12 AMI Trides Operator s Manual A 96 250 101...
Page 32: ...3 Structure and Function Software 24 AMI Trides Operator s Manual A 96 250 101...
Page 40: ...5 Diagnostics Menu Diagnostics 2 32 AMI Trides Operator s Manual A 96 250 101...
Page 64: ...6 Maintenance Repair Calibration 56 AMI Trides Operator s Manual A 96 250 101...
Page 76: ...7 Operation Logger 4 4 optional 68 AMI Trides Operator s Manual A 96 250 101...
Page 123: ...8 Installation Instrument Setup AMI Trides Operator s Manual A 96 250 101 115...
Page 128: ...9 Shut Down Transport Disposal Disposal 120 AMI Trides Operator s Manual A 96 250 101...