8 - Installation
Instrument Setup
The following screen sequences all show Relay 1. Note that the
Relay 2 sub-menus are identical to the ones of Relay 1.
Relay Contacts
Alarm Relay
Relay 1
Relay 2
Input
5.3.2
Enter
Function
Limit upper
Limit lower
Control upw.
Control dnw.
Timer
Function
Limit upper, lower
Relay 1
5.3.2.1
Parameter
recommended:
Parameter
DIS.
Setpoint
default for
Relay 2
is 10.00 ppm
Function
Limit upper
Parameter
DIS.
Setpoint
5.00 ppm
Hysteresis
0.10 ppm
Delay
30 Sec
DIS.
pH
Temperature
Sample flow
Relay 1
DIS
Relay 2
pH
Function
Limit upper, lower
Relay 1
5.3.2.20
Parameter
pH
Function
Limit upper
Parameter
pH
Setpoint
7.00 pH
Hysteresis
0.10 pH
Delay
30
Sec
Function
Limit upper, lower
Relay 1
5.3.2.20
Parameter
Temperature
Function
Limit upper
Parameter
Temperature
Setpoint
30.0 ° C
Hysteresis
1.0 ° C Delay
30
Sec
Function
Limit upper, lower
Relay 1
5.3.2.20
Parameter
Sample
flow
Function
Limit upper
Parameter
Sample flow
Setpoint
300 rpm
Hysteresis
30 rpm
Delay
30 Sec
AMI Trides Operator’s Manual, A-96.250.101
97
Summary of Contents for A-85.119.010
Page 10: ...1 AMI Trides keys 2 AMI Trides Operator s Manual A 96 250 101...
Page 15: ...2 Safety Instructions General 2 2 1 AMI Trides Operator s Manual A 96 250 101 7...
Page 20: ...3 3 1 2 AMI Trides pH 3 2 3 12 AMI Trides Operator s Manual A 96 250 101...
Page 32: ...3 Structure and Function Software 24 AMI Trides Operator s Manual A 96 250 101...
Page 40: ...5 Diagnostics Menu Diagnostics 2 32 AMI Trides Operator s Manual A 96 250 101...
Page 64: ...6 Maintenance Repair Calibration 56 AMI Trides Operator s Manual A 96 250 101...
Page 76: ...7 Operation Logger 4 4 optional 68 AMI Trides Operator s Manual A 96 250 101...
Page 123: ...8 Installation Instrument Setup AMI Trides Operator s Manual A 96 250 101 115...
Page 128: ...9 Shut Down Transport Disposal Disposal 120 AMI Trides Operator s Manual A 96 250 101...