SV 258 PRO (AG) User Manual
10
Note: Only SVAN 958A and the controller can be disconnected and removed from the station case
by the user. All other disassembling work should be performed strictly by an authorized service
team.
Note: The producer does not recommend removing the controller without a sound reason. Double
check that the controller has a good fixation in the socket after reconnecting!
The monitoring station case is equipped with an air pressure
compensation valve that enables easily opening of the case
if the internal pressure is lower than the atmospheric one.
Use the silica gel bags (optional accessory) for absorbing
any remaining moisture after closing the lid.
The valve should be released if opening the top lid is
problematic.
Note: The valve must be closed when the station is used outdoors, otherwise the case is not sealed
against moisture.
The station case is equipped with four Souriau UTO type connectors for:
•
input signals from the tri-axial piezoelectric
accelerometer (
INPUT 1-3
),
•
input signals from the microphone (
INPUT 4
),
•
powering station and charging the internal battery
(
DC SUPPLY
),
•
external
interface
for
digital
input/output
(
EXTERNAL INTERFACE
). In the standard station
configuration, this socket can be used for alarm
signals output (for example, SP 272
light and
sound alarm).
To connect a cable to a case socket, start by lining up the
key on the plug and socket, then lock the connector by
turning the ring clockwise (only the ring close to the socket
will rotate).
To disconnect a cable, push the connector towards the
station and turn it counterclockwise. New connectors require
more force so using a closed hand is more effective than
using only fingers.