Error diagnosis
Error
Possible cause
Error rectification
Elements on the front stay
dark, device does not
function
No mains voltage
Check the mains supply
The mains cable on the outlet or
device is not connected correctly
or at all
Check the mains cable
connection
Device not switched on
Switch on the device
Defective mains fuse
Device defect, service required
Internal power supply defective Device defect, service required
Elements on the front work
normally, no RF output
power
Defective accessory
Replace the accessory
Defective device
Turn the device off and on again.
Notify Service of any error
message that is displayed.
Auto-Start button is
flashing
“Auto-Start” has to be
acknowledged when loading a
program.
Press the Auto-Start button. No
error.
The button continues flashing
after it is acknowledged.
Disconnect and check the
connected accessory.
Accessory is in contact with
tissue or defective.
Even though the accessory was
removed, the button continues
flashing.
Operating the CURIS
®
can
continue in the STANDARD
operating mode, then contact
Service.
Display of an error
message, e.g. Err 42
See listing below
See listing below
Error messages and their meaning
The CURIS
®
RF generator conducts extensive functional tests during the power-on self-test.
All safety-related functions are also checked continuously during operation. In case of errors, the
RF energy output is stopped and the error is indicated by an error message.
NOTE
After an error message is displayed, the device can only be reset by turning it off and
back on again.
The messages where resolution by the user is possible are listed below. All other error messages
are listed in the service manual.
NOTE
Contact the manufacturer or specialist dealer if a message not described here is
displayed.