CURIS
®
Instructions for Use
17
2. Connect the plug of the connection cable for the neutral electrode to the connection socket
(30) again. If a monopolar operating mode is selected, the red warning lamp (31) stops flashing
now. If a segmented neutral electrode is used, it must be correctly applied to the patient for the
alarm to turn off. This warning lamp does not flash with bipolar operating modes.
3. Connect the connection cable with the monopolar electrode handle to the socket (28). Activate
the chosen current with the finger switch on the electrode handle or the foot switch. The
operating mode indicator lamps for MONOPOLAR (8) and BIPOLAR (15) have to light up
according to the current type that is selected, and the RF activation signal has to sound.
WARNING
If the RF-activation signal sounds without the foot switch or electrode handle
connected, then the device is defective. Do not operate the device! A technical
inspection is required.
If the RF activation signal sounds with the foot switch or electrode handle connected
without activating one of the control elements, one of these accessories is defective.
Continuing to operate this accessory is not permitted! It must be replaced.