CURIS
®
Instructions for Use
24
9
Technical information
9.1
Technical data, standards, certification
Mains supply
100-240 V; 50/60 Hz
Power consumption without RF output
approx. 50 VA
at max. output power
approx. 500 VA
Protection class
I
IP-Code
IP20
MPG classification
II b
LF and RF leakage
currents
according to IEC 60601-2-2
Type
CF; defibrillator safe
RF output
parameters:
MONOPOLAR current
type
Output
On load resistor
max. HF output
current
Nominal
frequency
Modulation
frequency
CUT 1 (cutting)
max. 100 W ± 20 %
on 600
1.10 A
4.00 MHz
-
CUT 2 (cutting)
max. 80 W
± 20 %
on 600
1.10 A
4.00 MHz
33 kHz
CONTACT (coag.)
max. 80 W
± 20 %
on 400
1.20 A
4.00 MHz
33 kHz
SOFTSPRAY (coag.)
max. 60 W
± 20 %
on 600
0.90 A
4.00 MHz
33 kHz
BIPOLAR current type
BICUT 1 (cutting)
max. 80 W
± 20 %
on 300
1.20 A
4.00 MHz
-
BICUT 2 (cutting)
max. 80 W
± 20 %
on 300
1.30 A
4.00 MHz
33 kHz
EXCISE (cutting)
max. 80 W
± 20 %
on 300
1.20 A
4.00 MHz
33 kHz
MACRO (coag.)
max. 80 W
± 20 %
on
50
2.20 A
4.00 MHz
-
PRECISE (coag.)
max. 50 W
± 20 %
on
50
1.70 A
4.00 MHz
33 kHz
RaVoR™ (coag.)
max. 40 W
± 20 %
on
50
1.60 A
4.00 MHz
33 kHz
Operating mode
Intermittent INT 10 s /30 s corresp. 25 % ED
Mains fuse
2 x T 3.15 AH 250 V G 5x20mm
Signal level
RF indicator: 40-65 dB(A) adjustable
Alarm: 65 dB(A)
Weight
approx. 5.0 kg
Dimensions
W x H x D 320 mm x 170 mm x 385 mm
Standards
IEC 60601-1: 2005 + Cor.:2006 + Cor.:2010 + A1:2012
IEC 60601-1-2: 2007
IEC 60601-2-2: 2017
Conforming to MDD 93/42/EEC
Environmental
conditions for
transportation and
storage
Ambient temperature
Relative humidity
Atmospheric pressure
-25 °C to +70 °C
10 % to 100 %
500 hPa to 1060 hPa
Environmental
conditions for
operation
Ambient temperature
Relative humidity
Atmospheric pressure
+10 °C to +40 °C
30 % to 75 %
700 hPa to 1060 hPa