XLINK 500/100 Operations & Maintenance Manual page # 31
The key settings for thermistors are:
Measurement Type: Analog
Analog type: A or B
Equation to compute temperature (Celsius):
Steinhart((10000)*X/(VREF-X), A, B, C) (see table below for A, B, C)
Note: The 10000 is the reference resistor value in ohms.
Slope: 1.8; Offset: 32 to convert to Fahrenheit.
Model
TYPE
A
B
C
44007
44034
YSI MIX B, 5K
0.0012829
0.0002365
9.1883E-08
44006
44031
YSI MIX H,
10K
0.0010295
0.0002391
1.568E-07
44005
44030
YSI MIX B, 3K
1.40510E-03
2.36900E-04
1.019E-07
44008
44032
YSI MIX H,
30K
9.35401E-04
2.21060E-04
1.2747E-07
44016
44036
5600-0025
5600-0030
YSI MIX B,
10K
1.12610E-03
2.34545E-04
8.6359E-08
Campbell Scientific’s CS205 & CS107 temperature probe requires different wiring than th
e
above since it has a reference resistor built into the probe. The wires are connected as indicated
below on the left, and the equation will use the built-in template for CS205/107. The image on
the right indicates the probe’s internal wiring for
reference.
•
Jumper C+ to Vref
•
Purple (AG) ----------------- C+
•
Red (HI) ---------------------- C-
•
Black (EX) -------------------- AGND
•
Clear (G) --------------------- Earth Ground