Meas Type
Wind
Wind Type
Wind Dir Analog
Analog Type
0-5V A
Warmup
0 sec
Averaging Time
00:00:10
Sampling Interval
1.0 sec
Use Equation
On
Equation
X/VREF*355 {converts to degrees}
Please note how some settings have been excluded from the table above. Settings such as
Measurement Interval
and
Time
are left to the user to decide. Also
Wind Averaging
can
be set to Vector or Scalar depending on the user’s needs.
Thermistor
This section describes how to physically connect the sensor and how to configure the
setup in MONITOR.
The sensor used in this example is a PS103J2 thermistor. This sensor is a 10k ohm type
and has two leads, so we need to add our own “completion” resistor (0.1% or better).
This completion resistor will help MONITOR to calculate the resistance of the thermistor
by measuring the voltage generated by two resistances (a known and unknown).
Since our thermistor is a 10k Ohm type, we will use a 10kohm completion resistor. The
equation
10000*X / (VREF-X)
Where the 10000 is our completion resistor in ohms and X
is the voltage reading.
will give us the resistance of the thermistor, which can then be placed into the Steinhart-
Hart equation to calculate temperature. Note in MONITOR, the Steinhart equation
already converts the retuned value from Kelvin to Celsius.
To wire it up, connect like so (NOTE: The lead numbers are not marked with numbers,
this is just a reference.)
Completion Resistor
MONITOR Connections
Lead 1
Vref
Lead 2
ANALOG 0-5V (A or B, same as thermistor).
Thermistor
Lead 1
ANALOG 0-5V (either A or B)
Lead 2
AGND
Thermistor Setup
Value
31