Specifications
Specifications
Specifications
Specifications
Patient population: Adult and pediatric patients (age 3 and above).
Method of measurement: Oscillometric
Initial inflation pressure: 1/- 20mmHg
Blood pressure range (mmHg): 60< Systolic BP< 270, 30< Diastolic BP< 170
Blood pressure accuracy: Measurements determined with this device are equivalent to those
obtained by a trained observer using the cuff/stethoscope auscultation method, within
the limits prescribed by the American National Standard,
Manual, electronic or
automated sphygmomanometers
.
Blood pressure determination time: 30-40 seconds typical for Adult cuff
Pulse rate range: 30-200 bpm +/- 2% or +/- 3 bpm, whichever is greater
Temperature range: 86
°
F (30.0
°
C) – 109.4
°
F (43.0
°
C)
Temperature accuracy: +/- 0.2
°
F (+/-0.1
°
C)
Functional oxygen saturation range: 40-100%
Functional oxygen saturation accuracy: 70-100% +/- 2 digits
(Note: because pulse oximeter equipment measurements are statistically distributed,
only about two-thirds of pulse oximeter equipment measurements can be expected to
fall within ± 2 digits of the value measured by a CO-oximeter.)
Operating conditions: 10
°
C (50
°
F) to 40
°
C (104
°
F) Less than 90% RH
Storage conditions: -20
°
C (-4
°
F) to 55
°
C (131
°
F) Less than 90% RH
Power: External power supply, AC adapter (
ADC
item number 9000AC9V)
Calibration: Check once per year for BP and Temperature