36
Replacing the Rechargeable Battery
Replace the battery:
•
According to your regular maintenance schedule.
•
When the battery no longer charges.
•
After heavy use, if necessary.
To replace the battery:
1.
Remove the four screws securing the battery bay door.
2.
Carefully remove the battery from the battery bay, being careful not to pull on
the wires attached to the battery terminals.
3.
Disconnect the wires from the battery terminals.
4.
The rechargeable battery contains lead. Please dispose of the old battery properly.
5.
Connect the wires to the terminals of the replacement battery, ensuring the red
wire is attached to the red terminal and the black wire to the black terminal. If
the wires are reversed, no damage will occur, however the
ADview
will not
operate. Be sure to use
ADC
item number 9000BAT for the replacement battery
in order to maintain optimum performance.
6.
Re-secure the battery bay door with the four screws removed in step 1.
7.
Connect the power supply to turn the device on and charge the replacement
battery fully before using.
If the rechargeable battery is disconnected for any reason, the device must
be connected to AC mains power via the power supply before the unit will
turn on; this is required even if the battery has been properly reconnected.
Disposal
This symbol indicates the device contains materials (such as electrical components)
which are hazardous. Please return to
ADC
for disposal.
Attaching the Temperature Module
Summary of Contents for ADC ADview 9000BP
Page 2: ...ii...
Page 80: ...69 IB p n 93 9000UM 00 rev 4...