TFG6800 Series Function/Arbitrary Waveform Generator Users’ Guide
Suin Instruments Co., Ltd
66
The rising edge of synchronous signal is corresponds to the starting point of each arbitrary
waveform in one sequence.
If external trigger or manual trigger is used, then synchronous signal will be determined by
external or manual trigger signal.
3.17.8 Sequence storage
The data of arbitrary sequence is stored in RAM memory and which will be lost once power off.
For long time storage, user needs to save the arbitrary wave sequence data to FLASH memory.
Press soft keys
〖
Store Recall
〗
and
〖
Save As File
〗
to enter the file management interface, press
〖
Open Dir
〗
soft key and then
〖
Options
〗
and
〖
New File
〗
to create a file named Arb_seq1.wseq
(you do not need to enter an extension, and the instrument automatically adds an extension when
creating a file,). Save arbitrary sequence data to a file, which is a binary-type file and multiple
such files could be saved and could be recall at any time for use.
3.17.9 Sequence Recall
After entering the operation interface of arbitrary wave sequence, press soft keys
〖
Store Recall
〗
and
〖
Recall File
〗
to enter the file management interfaces if arbitrary sequence file has been
stored previously, press
〖
Open Dir
〗
and find the sequence file Arb_seq1.wseq, press the soft key
〖
Options
〗
and
〖
WaveSeqRecall
〗
to load this sequence.
3.18 Frequency Counter
The waveform generator can measure the frequency and period of periodic continuous signal.
Besides, pulse width and duty cycle can also be measured for square and pulse.
Press
【
Counter
】
key to enter into counter menu.