Programmable DC Power Supply User’s Guide
28
For example: MEAS:VOLT CH1? (means to query CH1’s voltage value)
Return information: 1.00
Command format
:
MEASure:CURRent {CH1|CH2 }?
Function description: to query current value of specified channel’s output terminal
For example: MEAS:CURR CH1? (means to query CH1’s current value)
Return information: 1.000
Command format
:
MEASure:POWEr {CH1|CH2 }?
Function description: to query power value of specified channel’s output terminal
For example: MEAS:POWE CH1? (means to query CH1’s power value)
Return information: 0.15
Command format:
MEASure:OVP {CH1|CH2}?
Function description: to query OVP value of specified channel’s output terminal
For example: MEAS:OVP CH1?(means to query CH1’s OVP value)
Return information: 35.000
VOLTage
Command format:VOLTage {CH1|CH2|CH3}?
Function description: to query voltage value of specified channel.
Command format:
VOLTage {CH1|CH2,}<volt >
Function description: to set voltage value of present operation or specified channel.
For example: VOLT CH2,30.00 (means to set CH2’s voltage value to be 30.000V)