6541-4-810
Impaction Handle
6541-4-825
Slip Torque Handle
See Catalog
CR , CS & PS Tibial Inserts
Instrument Bar
41
Component Implantation
Tibial Insert Implantation
Figure 59
>
Prior to assembly of the Tibial Insert, the Tibial
Trial Insert may be placed on the Triathlon
Tritanium Baseplate, to once more assess joint
stability and range of motion .
>
To assemble the Tibial Insert, distract the joint
and angle the insert posteriorly into the Triathlon
Tritanium Baseplate . The posterior lip of the
Tibial Insert must fit beneath the lip on the
posterior Triathlon Tritanium Baseplate wall .
>
Attach the Tibial Insert Impactor to the
Impaction Handle and impact to snap the Insert
in place anteriorly .
>
The Tibial Insert is fully seated once the locking
wire locks under the barbs on the anterior/
interior surface of the Triathlon Tritanium
Baseplate wall .
6541-4-805
Baseplate Impactor/Extractor
6541-4-813
Tibial Insert Impactor
Size 1-3 – 6541-2-013
Size 4-6 – 6541-2-046
Cemented Keel Punch